Operating Instructions
TDR sensor for continuous level and
interface measurement of liquids
VEGAFLEX 86
Modbus and Levelmaster protocol
Document ID: 41848
Przeglądanie stron 0
1 2 3 4 5 6 ... 103 104

Podsumowanie treści

Strona 1 - VEGAFLEX 86

Operating InstructionsTDR sensor for continuous level and interface measurement of liquidsVEGAFLEX 86Modbus and Levelmaster protocolDocument ID: 41848

Strona 2 - Quick start

103 Product descriptionVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-1303143 Product description3.1 CongurationThe nameplate contains the m

Strona 3

10011 SupplementVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-13031411.8 Industrial property rightsVEGA product lines are global protected by

Strona 4

101INDEXVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314INDEXAAccessories – Display and adjustment module 14 – External display and adjus

Strona 5 - Contents

102INDEXVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314 – Bypass 16 – Display and adjustment module with heating 15 – Electronics modul

Strona 6

103NotesVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314

Strona 7 - 1 About this document

Printing date:VEGA Grieshaber KGAm Hohenstein 11377761 SchiltachGermany41848-EN-130314All statements concerning scope of delivery, application, practi

Strona 8 - 2 For your safety

113 Product descriptionVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-13031421Fig. 5: Position of Modbus and sensor electronics1 Modbus electr

Strona 9 - 2.6 NAMUR recommendations

123 Product descriptionVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314dh1Fig. 6: Level measurement1 Sensorreferenceplane(sealsurface

Strona 10 - 3 Product description

133 Product descriptionVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314TSd1h1h2d21L2L1L3Fig. 7: Interface measurement1 Sensorreferencep

Strona 11 - 3.2 Principle of operation

143 Product descriptionVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314For the interface measurement, you can select the requested output

Strona 12 - 3 Product description

153 Product descriptionVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314VEGADIS 61 is an external display and adjustment unit for sensors w

Strona 13

163 Product descriptionVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314If you are using an instrument with rod version, you can extend the

Strona 14

174 MountingVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-1303144 Mounting4.1 General instructionsWith instruments with threaded process tti

Strona 15

184 MountingVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314sible. The exact value of the min. distance (lower dead band) is stated in cha

Strona 16

194 MountingVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314If possible, avoid sockets. Mount the sensor ush with the vessel top. If this

Strona 17 - 4 Mounting

2Quick startVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314Quick startThe quick start procedure enables a quick setup with many applica-t

Strona 18

204 MountingVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314Fig.12:MountingofthesensorwithinowingmediumThe reference plane for th

Strona 19

214 MountingVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314Generally a spacer at the probe end is sucient for rod probes to avoid contac

Strona 20

224 MountingVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314Information:The distance piece must only be incorporated up to max. 50 mm (2 i

Strona 21

234 MountingVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314You can nd further information in the supplementary instructions of the rod e

Strona 22

245 Connecting to power supply and bus systemVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-1303145 Connecting to power supply and bus system5

Strona 23

255 Connecting to power supply and bus systemVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-1303145.2 ConnectionThe voltage supply and signal o

Strona 24 - 5.1 Preparing the connection

265 Connecting to power supply and bus systemVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-1303148. Connect the screen to the internal ground

Strona 25 - 5.2 Connection

275 Connecting to power supply and bus systemVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314++power supplyMODBUSD0D1IS GNDUSB1213452off

Strona 26 - 4...20mA

286 Set up with the display and adjustment moduleVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-1303146 Set up with the display and adjustment

Strona 27

296 Set up with the display and adjustment moduleVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-1303146.3 Adjustment system12Fig. 21: Display

Strona 28 - 6.1 Adjustment volume

3Quick startVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-1303141. In this menu item you can select the application. You can choose between l

Strona 29 - 6.3 Adjustment system

306 Set up with the display and adjustment moduleVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314• Software version (SW-Ver)• Hardware ver

Strona 30

316 Set up with the display and adjustment moduleVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314Enter the appropriate distance value in m

Strona 31

326 Set up with the display and adjustment moduleVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314LinearizationA linearization is necessary

Strona 32

336 Set up with the display and adjustment moduleVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314Info: Instrument name, hardware and softw

Strona 33

346 Set up with the display and adjustment moduleVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314In this menu item, you can select the app

Strona 34

356 Set up with the display and adjustment moduleVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314This menu item is only available if you h

Strona 35

366 Set up with the display and adjustment moduleVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314centage value. The distance refers tot he

Strona 36

376 Set up with the display and adjustment moduleVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314If you have selected interface measuremen

Strona 37

386 Set up with the display and adjustment moduleVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314In menu item "Level 2" you den

Strona 38

396 Set up with the display and adjustment moduleVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314In menu item "Current output, adjust

Strona 39

4Quick startVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314other distances, for example, 10 % and 90 % is also possible. Starting point f

Strona 40

406 Set up with the display and adjustment moduleVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314The following circumstances cause interfe

Strona 41

416 Set up with the display and adjustment moduleVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314In the menu item "Lock/unlock adjust

Strona 42

426 Set up with the display and adjustment moduleVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314The optionally integrated background ligh

Strona 43

436 Set up with the display and adjustment moduleVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314In another window you can carry out a res

Strona 44

446 Set up with the display and adjustment moduleVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314Select the requested simulation size and

Strona 45

456 Set up with the display and adjustment moduleVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314In delivery status, the PIN is "0000

Strona 46

466 Set up with the display and adjustment moduleVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314Menu Menu item Default valueSetup Block o

Strona 47

476 Set up with the display and adjustment moduleVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314The copied data are permanently saved in

Strona 48

486 Set up with the display and adjustment moduleVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314The default setting is "Analogue cur

Strona 49

496 Set up with the display and adjustment moduleVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314• All data of the menu "Setup"

Strona 50 - 7.1 Connect the PC

5ContentsVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314Contents1 About this document1.1 Function ...

Strona 51

507 Setting up sensor and Modbus interface with PACTwareVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-1303147 Setting up sensor and Modbus in

Strona 52

517 Setting up sensor and Modbus interface with PACTwareVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314The connection of the PC to the RS

Strona 53 - 3 Maintenance

527 Setting up sensor and Modbus interface with PACTwareVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314Fig. 25: Example of a DTM viewAll

Strona 54

537 Setting up sensor and Modbus interface with PACTwareVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314Fig. 26: Select quick setup1 Quic

Strona 55

547 Setting up sensor and Modbus interface with PACTwareVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314Device nameHere, you can nd the i

Strona 56

557 Setting up sensor and Modbus interface with PACTwareVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314Interface measurement: If you sele

Strona 57

567 Setting up sensor and Modbus interface with PACTwareVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314• Linear• Spherical tank• Horizont

Strona 58 - 8 Set up with other systems

577 Setting up sensor and Modbus interface with PACTwareVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314LockadjustmentwithPINafterset

Strona 59 - 9 Diagnostics and service

588 Set up with other systemsVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-1303148 Set up with other systems8.1 DD adjustment programsDevice

Strona 60 - 9.3 Status messages

599 Diagnostics and serviceVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-1303149 Diagnostics and service9.1 MaintenanceIf the device is used

Strona 61 - 9 Diagnostics and service

6ContentsVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-1303149.5 Exchanging the electronics module ... 669.6

Strona 62

609 Diagnostics and serviceVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314• Display and adjustment moduleFurther echo curves: Up to 10 ec

Strona 63

619 Diagnostics and serviceVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314This status message is inactive by default. It can be activated

Strona 64 - 9.4 Rectify faults

629 Diagnostics and serviceVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314CodeText mes-sageCause Rectication PA DevS-pecDiagnosisF125Unp

Strona 65

639 Diagnostics and serviceVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314CodeText mes-sageCause RecticationS600Unpermissi-ble electroni

Strona 66

649 Diagnostics and serviceVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-1303149.4 Rectify faultsThe operator of the system is responsible fo

Strona 67

659 Diagnostics and serviceVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-13031412Leveltime0Fig. 28: The broken line 1 shows the real level, th

Strona 68

669 Diagnostics and serviceVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314Fault description Error pattern Cause Rectication5. Measured v

Strona 69

679 Diagnostics and serviceVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314ics modules are adapted to the respective sensor and dier in s

Strona 70 - 10 Dismounting

689 Diagnostics and serviceVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314Information:Please keep the stated torque so that the max. tens

Strona 71 - 11 Supplement

699 Diagnostics and serviceVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314• Interface adapter VEGACONNECT 4• PC with PACTware• Current se

Strona 72

71 About this documentVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-1303141 About this document1.1 FunctionThis operating instructions manual

Strona 73

7010 DismountingVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-13031410 Dismounting10.1 Dismounting stepsWarning:Before dismounting, be aware

Strona 74

7111 SupplementVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-13031411 Supplement11.1 Technical dataGeneral data316L corresponds to 1.4404 or

Strona 75

7211 SupplementVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314 Ʋ Cable: ø 2 mm (0.079 in) approx. 16 g/m (0.17 oz/ft) Ʋ Cable: ø 4 mm (0.

Strona 76

7311 SupplementVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-1303142435124531Fig. 39: Measuring ranges - VEGAFLEX 861 Reference plane2 Probe

Strona 77

7411 SupplementVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314 Ʋ Temperature +18 … +30 °C (+64 … +86 °F) Ʋ Relative humidity 45 … 75 % Ʋ

Strona 78

7511 SupplementVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314L2mm(0.079")-2mm(0.079")025mm(0.984")-25mm(-0.984")

Strona 79

7611 SupplementVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314L0-15mm(-0.591")2mm(0.079")-2mm(-0.079")25mm(0.984"

Strona 80 - 11.2 Basics Modbus

7711 SupplementVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314The following table shows the resulting deviation for some typical gases an

Strona 81 - Protocol description

7811 SupplementVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314120°C(32°F)-40°C(-104°F)65°C(149°F)80°C(176°F)280°C(536°F)250°C(482°F)190°C

Strona 82 - 11.3 Modbus register

7911 SupplementVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-13031421-100°C(-148°F)-200°C(-328°F)100 bar(1450 psig)200 bar(2900 psig)300 bar(4

Strona 83

82 For your safetyVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-1303142 For your safety2.1 Authorised personnelAll operations described in t

Strona 84 - 11.4 Modbus RTU commands

8011 SupplementVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314Display and adjustment moduleDisplay element Display with backlightMeasured

Strona 85 - FC8 Diagnostics

8111 SupplementVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314data transfer. In the RTU and ASCII mode interesting here, data will be tra

Strona 86 - FC17 Report Slave ID

8211 SupplementVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314Parameter CongurableValues Default ValueData Bits 7, 8 8Parity None, Odd,

Strona 87 - 11.5 Levelmaster commands

8311 SupplementVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314Register Name Register Number Type NoteStatus 100 DWord Bit 0: Invalid Meas

Strona 88 - Report Unit Number

8411 SupplementVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314Register Name Register Number Type NoteTV 2106 DWord Third Variable in Byte

Strona 89 - Set Baud Rate

8511 SupplementVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314Parameter Length Code/DataNumber of Registers 2 Bytes 1 to 127 (0x7D)Respon

Strona 90 - Error codes

8611 SupplementVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314Request:Parameter Length Code/DataFunction Code 1 Byte 0x08Sub Function Cod

Strona 91 - Fisher ROC 809

8711 SupplementVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314Response:Parameter Length Code/DataFunction Code 1 Byte 0x11Byte Number 1 B

Strona 92 - ABB Total Flow

8811 SupplementVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314Parameter CongurableValues Default ValueBaud Rate 1200, 2400, 4800, 9600,

Strona 93 - Thermo Electron Autopilot

8911 SupplementVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314Response:Parameter Length Code/DataReport Level (and Temperature) 6 charact

Strona 94 - Bristol ControlWave Micro

92 For your safetyVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314Electromagnetic compatibilityInstruments with plastic housing as well as

Strona 95 - ScadaPack

9011 SupplementVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314Set Receive to Transmit DelayRequest:Parameter Length Code/DataSet Receive

Strona 96

9111 SupplementVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-13031411.6 CongurationoftypicalModbushostsFisher ROC 809Wiring plan++power

Strona 97

9211 SupplementVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314ABB Total FlowWiring plan++power supplyMODBUSD0D1IS GNDUSB1 213452offon(

Strona 98

9311 SupplementVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314Thermo Electron AutopilotWiring plan++power supplyMODBUSD0D1IS GNDUSB1231

Strona 99 - VEGAFLEX 86, rod version

9411 SupplementVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314Bristol ControlWave MicroWiring plan++power supplyMODBUSD0D1IS GNDUSB1231

Strona 100 - 11.9 Trademark

9511 SupplementVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314ScadaPackWiring plan++power supplyMODBUSD0D1IS GNDUSB13213455213496782off

Strona 101

9611 SupplementVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314Housing1~ 87 mm (3.43")M16x1,5ø 84 mm(3.31")120 mm (4.72")M2

Strona 102

9711 SupplementVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314VEGAFLEX 86, cable version with gravity weight412 3SW 60(2.36")G1½ /1½

Strona 103 - 41848-EN-130314

9811 SupplementVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314VEGAFLEX 86, cable version with centering weightG1½ /1½ NPTL35mm(1.38"

Strona 104 - Printing date:

9911 SupplementVEGAFLEX 86 • Modbus and Levelmaster protocol41848-EN-130314VEGAFLEX 86, rod versionSW 60(2.36")G1½ /1½ NPTø 16mm(0.63")L35mm

Komentarze do niniejszej Instrukcji

Brak uwag