3 Product description
VEGAFLEX 83 • Modbus and Levelmaster protocol
41838-EN-130314
TS
d1
h1
h2
d2
1
L2
L1
L3
Fig. 7: Interface measurement
1 Sensorreferenceplane(sealsurfaceoftheprocesstting)
d1 Distancetotheinterface(HARTvalue1)
d2 Distancetothelevel(HARTvalue3)
TSThicknessoftheuppermedium(d1-d2)
h1 Height - Interface
h2 Height - Level
L1 Lower medium
L2 Upper medium
L3 Gas phase
Upper medium (L2)
The upper medium must not be conductive
The dielectric constant of the upper medium must be known (input
required). Min. dielectric constant: rod version 1.6. You can nd a
list of dielectric constants on our home page: www.vega.com
The composition of the upper medium must be stable, no varying
products or mixtures
The upper medium must be homogeneous, no stratications
within the medium
Min. thickness of the upper medium 50 mm (1.97 in)
Clear separation from the lower medium, no emulsion phase, no
mull layer
If possible, no foam on the surface
Lower medium (L1)
The dielectric constant must be 10 higher than the dielectric
constant of the upper medium, preferably electrically conductive.
Example: upper medium dielectric constant 2, lower medium at
least dielectric constant 12.
Gas phase (L3)
Air or gas mixture
Gas phase - dependent on the application, gas pahse does not
always exist (d2 = 0)
The instrument is always preset to the application "Level measure-
ment".
Prerequisites for inter-
face measurement
Output signal