7 Setting up sensor and Modbus interface with PACTware
VEGAFLEX 83 • Modbus and Levelmaster protocol
41838-EN-130314
Linear
Spherical tank
Horizontal cylindrical tank
You must enter the following vessel dimensions with non-linear condi-
tions:
Height of the socket h
Vessel height D
In this window you can adjust the output signal. When the function is
inactive, you can change the settings via the "Extended adjustment".
These settings allow you to optimize the sensor. With them you can
compare a sounded distance with the indicated value and correct it, if
necessary.
Probe immersed in the liquid (covered)
Select whether or not the probe is immersed in the medium.
Measured distance to the medium
If the probe is immersed in the medium, you can enter here the meas-
ured distance to the medium.
Displayeddistancecorrect?
Is the displayed distance value correct? If you have the possibility, you
can enter here the sounded distance to the medium.
False signal suppression
With this function you can carry out the automatic false signal sup-
pression. We recommend carrying out false signal suppression in any
case.
When the setup of the instrument is nished, additional settings can
be made. These are various backups and the locking of the instru-
ment against unauthorised or inadvertent adjustment.
Prepareabackupleoftheinstrumentparameteradjustment?
For backup purposes, the current parameter adjustment of the instru-
ment is stored in a le. You can use this le later on to restore the
instrument parameter adjustment. The complete data set is down-
loaded from the device. This procedure can last several minutes.
Createinstrumentdocumentation?
This function is used to print or create a PDF le of the current param-
eter adjustment. To read the PDF le, you need a suitable program
(for example Acrobat Reader). To print or create the PDF le, all data
are downloaded from the device. For this function, the full version
of the DTM Collection is required. This procedure can last several
minutes.
Storeechocurveofthesetupinthesensor?
Have you completed the initial setup of the instrument? In such case,
we recommend storing the current signal conditions in the device for
later instrument tests and diagnostics.
Step 5 of 7
Setting - Output signal
Step 6 of 7
Sensor optimization
Step 7 of 7
Additional settings